Wavetek 39A Specifikace Strana 113

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 137
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 112
User Manual 301/302
TRIGger:SOURce:ADVance EXTernal | INTernal
Purpose
This command will set the advance source for the trigger mode.
<EXTernal> will select the external input
<INTernal > will select the internal trigger generator
TRIGger:SOURce:ADVance?
Response
The 302 will return EXT, or INT depending on the present trigger source setting.
TRIGger:TIMer <period>
Purpose
This command will set the period for the internal trigger generator.
<period> numeric value will set the frequency of the internal trigger generator
TRIGger:TIMer?
Response
The 302 will return the present internal period value.
TRIGger:IMMediate
Purpose
This command will stimulate the 302 output once. This command will affect the 302 only after it will be set
to interrupted operation.
*TRG
Purpose
This common command has the same affect as the TRIG:IMM command.
SCPI Command Reference
102
Zobrazit stránku 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 136 137

Komentáře k této Příručce

Žádné komentáře